If Nebulisation performance Is reduced!

If you have noticed a reduction in the performance of your Omron MicroAIR and Medix MicroNeb nebuliser and have cleaned as instructed under
How to clean and How to disinfect then check the power supply.
*Check batteries
*Use the AC direct current adapter [if you have one]
It may be your Mesh holes in the mesh cap have become clogged.

What to do about clogged holes In the mesh cap!

NE-U22 is designed to nebulise at least 2ml of saline for 8 minutes, If it takes more than 8 minutes to
nebulise 2m1 of saline, it is possible that the holes In the mesh are clogged. If saliva has stained the
mesh the nebulisation rate may decrease. If the stain cannot be removed by either the daily maintenance process or boiling
it may be necessary to follow the procedure below to recover the performance of the mesh.
•Boil the mesh cap for ten minuets in a mixture of water and white wine vinegar at a strength of 3 parts water one part
vinegar, 3:1
After boiling remove from hot water and rinse In a bowl of cold water. Leave to air dry on a lint free cloth or surface,
If the performance of, your MicroAIR  or MicroNeb is not restored to your expectations use the following cleaning procedure:
Sometimes reduced performance can be due to saliva coming into contact with the mesh.
° Saliva contains protein and it is recommended to use a detergent which contains "protein removing properties" such as
'false teeth detergent~.
•There are sever kinds of detergent available in your market It is recommended to use a cleaning agent that has neutral or
weak alkaline chemical characteristic.
•Be sure to select a denture cleaner, which is a liquid solution and not a salt-based suspension. (Ask your Pharmacist for
help),

How to clean the mesh with denture cleaner

1, To make denture cleaner detergent solution according to instructions of the manufacturer.
2, Soak the mesh cap for 8-10 hours,
3. To remove the denture cleaner detergent left in the mesh, rinse the mesh cap in water and then boil the cap for 20
minutes.
4 Wipe off water with new gauze and dry them thoroughly.
(Caution, Do not touch parts while they are hot from boiling-)

Dish washing machines can be used but this is not the preferred method for the cleaning of the MicroAIR or MicoNeb. The MicroAIR and MicroNeb has
some small parts that may possibly be flung around inside a machine by the water Jets that could damage the accessories.
A dishwasher can be used to wash the drug chamber, mesh cap, masks, mouth piece and cover but not the main body of the unit.
It may be necessary to rinse the mesh cap after a dishwashing depending on any salt residue that may
remain on the mesh from washing.

If after following these procedure, you are not satisfied with the performance of your MicroAIR, please contact our customer
Service, after-sales desk. The Local number Is on the box, and an the instruction manual supplied with this MicroAIR,

After following the Instructions above
My nebuliser mesh on my MicroAIR or MicroNeb continues to under perform, then you may need to replace the mesh. With average use
(1-4 times a day for two minutes) the mesh should last about a year. Please order a replacement mesh for your MicroAIR or MicroNeb so
you can return to using you nebuliser to its optimum performance.